Ultra Installation Guide v2.0
38
Customer Support
1.800.366.5343 –
VZ0022UG201509-2.0
Generator Auto Calibration:
1.
Select “Auto Calibration” button.
2.
Select workstation (WS3 / Direct mode)
3.
Angle Tube away from detector
4.
Click “OK”.
5.
Confirm or leave the Auto-calibration by pressing the respective button (“Yes” or “No”) on the
new screen.
6.
Select focal spot calibration. Figure 23 (left)
7.
Click “OK”.
8.
Select “Yes” Figure 23 (right)
Figure 2
1.
Confirm or leave the Auto-calibration (second confirmation) by pressing the respective button
(“Confirm” or “Dismiss”) on the calibration screen. Figure 24
2.
Auto-calibration is activated when the “AutoCalibration” is shown. Figure 25