Ultra Installation Guide v2.0
41
Customer Support
1.800.366.5343 –
VZ0022UG201509-2.0
Positioner Calibration
c.
Change Units from CM to Inch
d.
Set Safety Distance for 3.9”, Press SET
e.
Room Height to verified Room Height, Press SET
f.
Detector Width Left to 10.6”, Press SET
g.
Detector Width Right to 16.1”, Press SET
h.
Detector Thickness to 5.9”, Press SET
Height Calibration
3.9”
Set Height
10.6”
16.1”
5.9”
Rotate the arm to the horizontal position. Verify
that the arm angle is 90 degrees with a digital
level placed on the SID portion of the column.
Position the Arm at corresponding calibration
Spot.
Measure with Metric Tape from the Arm
Central guide black dot to the floor
Lower ramp on back of column to calibrate the
detector lower to the floor and change MIN
HEIGHT to match calibrated height