Ultra Installation Guide v2.0
50
Customer Support
1.800.366.5343 –
VZ0022UG201509-2.0
a.
3000 +/- 500; ~ (80kvp 32mAs 160ma 200ms)
b.
5000 +/- 500; ~ (80kvp 50mAs 160ma 320ms)
c.
6000 +/- 500; ~ (80kvp 64mAs 160ma 400ms)
23.
Select Complete
24.
Select Finish
25.
Select To standby
26.
Select Disconnect
27.
Exit TETD_FPD_Manager64.exe
28.
Launch Ultra
•
On first launch of Ultra after calibration let sit on work list screen for one minute.
29.
Launch Aquire. Verify Configured or idle Status. Change MS to 640. If Status Changes from configured to
switching, and then back to configured. Then Calibrations copied over correctly. If not
a.
Exit Ultra
b.
Open C:\opal\plugins64\Toshiba\”1_serial Number”
c.
Copy Defect bin, DefectMap,Gain High, Gain Low, Fain Mid, Gain Off High,Gain Off Mid, Gain
Tumb Min, Paste these into “C:\opal\plugins64\Toshiba\”2_SerialNumber”’.
d.
Open C:\opal\plugins64\Toshiba\”1_serial Number”\OffsetLUT
e.
Copy Offse A and Offset C
f.
Paste into C:\opal\plugins64\Toshiba\”2_serial Number”\OffsetLUT
g.
Launch Ultra. Verify with test Shots.