VET-PRO
Operator’s Manual
7
•
Although the pump will operate using a variety of giving sets, use of
other administration sets may impair the operation of the pump and the
accuracy of the infusion.
F
Note:
The maximum volume that may be infused under
SINGLE FAULT
CONDITION
is 0.1 ml
G
CAUTION
To avoid possible damage to the equipment, observe the
following:
•
Do not immerse the pump in liquids.
•
Clean solution spills on the pump immediately. Use a damp
cloth or sponge. A cleaning disinfectant may be used. Wipe
thoroughly with a dry cloth.
•
Do not clean pump with chemicals such as Xylene, Acetone or
similar solvents. These chemicals cause damage to plastic
components and paint.
I
Warning:
Any adjustments, maintenance, or repair of the uncovered
pump while connected to the power should be avoided. If
necessary, these repairs should be carried out by skilled
technicians who are aware of the hazards involved.
I
Warning:
The pump should be operated within an ambient temperature
range of 15
°
C to 35
°
C. and at up to 85% humidity
I
Warning
:
Do not operate the pump near high-energy radio-frequency
emitting equipment, such as electro-surgical cauterizing
equipment, and cellular telephones. False alarm signals may
occur.