VET-PRO
Operator’s Manual
21
Door Open - Alarm Test:
1.
Insert an administration set into the pump (as described in page 17) and flush
it (as described on page 18).
Rate 500 ml/h
2.
Set the infusion rate at 500ml/hr, using the
numeric keyboard.
Press OK for Volume
Volume 20 ml
3.
Set Volume at 20 ml.
Press OK to start
3.
Press the
START/OK
key to start operation.
4. Open the pump door.
Door Open
5
Door Open alarm will be activated at once.
Close door
6.
Close the door.
STOP
2 min
7.
STOP appears on the display.
Press OK to start