VET-PRO
Operator’s Manual
50
From STOP mode, switch the pump off by
pressing the
ON/OFF
key (continuously
until a beep is sounded)
.
OFF
WAIT FOR BEEP
F
Note:
The pump cannot be switched off during operation.
Important
: Use the technique dictated by the IV container provider.
1. Ensure that the pump is not infusing by placing the pump in STOP mode or
turning the pump OFF.
2. Remove the bag spike on the IV set from the empty / used container. Discard
the empty/used container according to the providers instructions.
3. Prepare a new container as per the providers instructions and insert the spike
into the new container.
4. Squeeze the drip chamber (if applicable) to fill approximately halfway with
fluid.
5. Restart infusion.
Locking
The
VET-PRO
enables the user to lock the programs, on three different
levels.
Level 1 (keys locking)
Level 2 (lock out program)
Level 3 (presetting maximal rates)
Level 1 – key locking
The
VET-PRO
keys locking function disables all but the
info
key,
START/OK
(during setting mode) and
STOP/NO
(during operation mode)
keys.
Level – program locking