VET-PRO
Operator’s Manual
58
Troubleshooting
Description
Result
Possible Cause
Remedy
Program completed
End PROGRAM
Press STOP to mute
The program was completed
KVO 5 ml/h
Press STOP to renew
Pump
will turn
to KVO
mode
KVO (Keep Vein Open)
program starts.
Press the
STOP/NO
key to restart
Show Setting
The program was designed to disable unauthourised user access to
technician settings.
Access to show setting mode, enable the user to
inspect all settings,
but not change them
.
Press
info
key
twice
.
Volume Infused
=>Show Setting
Choose "Show Setting" to display all software
adjustable parameters.
Adjust setting
Show Setting:
=>Battery Level
Scroll down the followung parameters on the
Show Setting mode:
Lock Out
Show Setting Options
Screen no.
Setting
Screen
1
Battery Level (Empty ? Full)
Battery Level
2
Lock Out Option (ON/OFF)
Lock Out
3
Time & Date display
Time, Date
4
Tube Temperature
Tube Temperature
5
Buzzer Level
Buzzer Level
6
Pressure level
(High/Normal/Low)
Pressure level
7
Pressure Default
Pressure Default
8
Select Program
Select Program
9
Display Flushing Volume (4 ml.
Flushing Volume