VET-PRO
Operator’s Manual
41
Bag Vol: 500 ml
Volume Left 500
Infused 0
17.
P
A
Press OK to start
The pump will display the protocol
accumultion screen. Use the arrow keys to
scroll through the data. Press the
START/OK
to start the program operation. Pressing the
STOP/NO
twice will go back to menu screen.
Note
: -If one of the values is changed during the programing mode,
the pump will not store the new value, unless the change is
confirmed by pressing the
START/OK
key.
MORPHINE
Rate 50 ml/h
18.
During the PCA program the display will show
the selected protocol, the name of the drug, the
rate of operating and the bolus volume.
P
A
BOLUS 5 ml
Volume 5 ml
19.
B
O
L
U
0 2.1 5
Pressing the Syringe key and approving with
START/OK
key will activate the
Bolus
Operation
. When the Bolus is given, the pump
will display the progress of the procedure.
A bolus can also be given by pressing twice
on the bolus cable.
20.
P
C
A
Resume:Press OK
Menu:Press NO
After a program has been locked, whenever the
pump will be turn on, resume-menu screen will
be displayed. Pressing
START/OK
key will
display accumulation screen, with Volume to
be infused, as volume left from previous
program. User can only continue previous
program by pressing again
START/OK
key.
Pressing
STOP/NO
key, will display code
setting screen. Set code 007, to enter the PCA
menu.
Start new bag
1.
Selecting “
Start new Bag”,
from the PCA
menu, will display another screen for approval.
P
C
A
History
=>
Start New Bag
Select Program