Program Setting
F
Note:
Whenever you decide to set a new program, follow
instructions at the bottom of screen.
Rate over Volume
Rate 500 ml/h
1.
The pump will display the last set rate.
Use the
numeric keys
to change or
press
START/OK
to confirm.
Press OK for Volume
Rate: 500 ml/h
During operation the pump allows
rate titration, by pressing a
numeric or digital point key.
Adjust & press OK
F
Note:
The rate on the display will change during titration, however
operation rate will change only if the
START/OK
key is
pressed within 10 seconds confirming the rate change. If the
rate change is not confirmed within 10 seconds, the pump
will continue operation on the original rate, and the display
will return to the previous rate set.
F
Note:
From the STOP mode, the user may return to the Rate screen
by pressing twice on one of the numeric keys.
F
Note:
During stop or data setting mode, use the
STOP/NO
key to
return one step backwards: First press will erase the last digit
displayed, once all the digits are erased, pressing the
STOP/NO
key will display the previous screen .
Volume 456 ml
Press OK to start
2.
When the Rate is confirmed, the
pump will display the last set volume
(if the time option is set off). Use the
numeric keys to change the
Volume
or press
START/OK
to confirm and
start the program operation.
Setting Program Delay– Refer to
page 48.