VET-PRO
Operator’s Manual
59
Is the default- up to 20 ml)
10
Showing Flushing Rate
(up to 1200 ml/hr)
Flushing Rate
11
Display BOLUS rate
Bolus Rate
Maximal Rate setting
Maximal Rate
Display KVO rate
KVO Rate
Air bubbles size
Air Size
Store last program
Store last Program
Operation led option (on/off)
Operation Led
Time option (on/off)
Time Option
18
Delay before start option (on/off)
Delay Before Start
19
Language
Language
20
Pump Serial Number
Serial Number
21
Production Date
Production Date
22
Operation time from service
(hours)
Operation Time
23
Software Version
Software Version
24
Volume Calibration
Volume Calibration
25
Pressure calibration - delta
Pressure Delta
26
Pressure calibration CAP
Pressure CAP
27
EXIT
EXIT
Adjust Setting
Press
info
key
twice
.
Volume Infused
Show Setting
Choose "Adjust Setting" to display all software
adjustable parameters.
=>Adjust Setting
Confirm with
START/OK
key.
Adjust Setting:
=> Pressure Level
See list of parameters that can be set on "Adjust
Setting" selection on the next page.
Buzzer Level
The first 2 parameters, Pressure level and Buzzer level, can be adjusted using
the arrow keys- no secret code required.
Buzzer level
=> More …
Access to “More…” will required a code.
Use
ARROW
keys to scroll to "More".
Show Setting