Changing Current Program:
1.
Switch the pump on by pressing
ON/OFF
key.
2.
When the screen of Rate appears, press
info
key twice.
Volume Infused
Show Setting
3.
Use
ARROW
keys to scroll to "Adjust Setting"
option.
=> Adjust Setting
4.
Choose "Adjust Setting" with
START/OK
key.
Buzzer level
=> More …
5.
Use
ARROW
keys (
2?, 0?)
to scroll to "More".
Show Setting
6.
Confirm with
START/OK
key.
Technician
7.
Set technician Code (Code 901), using the
numeric keyboard and confirm with
START/OK
key.
Code
901
Pressure Default
=> Select Program
8.
Scroll through the presetting parameters, using
the
ARROW
keys
(2?, 0?),
until the "Select
Program " option. Confirm with
START/OK
key.
Flushing Volume
Select program:
=>
CONTINUOUS
9.
Choose the required program, using the
ARROW
keys, and confirm with the
START/OK
key.
TPN
History
=> Restart Pump
10.
Choose “Restart Pump” Option and confirm
with the
START/OK
key.
Pressure Default
F
Note:
The selected program will remain in the pump’s memory until
changed by the user.