Cleaning
56
8.3.2
Rinsing
with
Purifying
Agent
After
pre
‐
purification
by
compressed
air,
residual
liquid
may
still
be
present
inside
the
fluid
system
(particularly
when
dispensing
high
‐
viscous
substances)
.
You
can
only
remove
it
with
a
flushing
procedure.
For
this
purpose,
the
following
agents
may
be
suitable:
Distilled
water
Ethanol
Isopropanol
(IPA)
Acetone
WARNING!
(potential
chemical
reactions)
Before
the
start
of
the
procedure,
carefully
read
the
safety
data
sheet
of
the
dis
‐
pensed
substance.
Ensure
compatibility
with
the
cleaning
agent.
The
cleaning
agent
itself
should
not
react
with
any
part
in
contact
with
the
fluid.
Consult
the
manufactur
‐
er
in
case
of
need,
especially
if
aggressive
material
is
intended
to
be
used.
In
case
of
need,
consult
the
manufacturer.
Step
1:
The
dispensing
process
must
be
completed.
The
valve
is
in
closed
position.
Do
NOT
switch
OFF
the
control
unit.
Step
2:
Separate
the
compressed
‐
air
connection.
The
pressure
must
be
reduced
to
0
bar.
Disconnect
the
PP
adapter
fitting
from
the
cartridge.
Step
3:
Replace
the
used
cartridge
by
a
new
one.
The
purifying
agent
can
be
supplied.
CAUTION!
All
of
the
connectors
must
be
installed
and
leak
‐
tight.
Step
4:
Reinstall
the
compressed
‐
air
connection
as
follows:
Place
the
PP
adapter
fitting
on
the
cartridge.
Rotate
clockwise,
until
it
latches
in
place.
Connect
the
PVC
hose
with
the
KS4
‐
CK
‐
6
coupler
plug
to
a
compressed
‐
air
supply.
For
this
purpose
you
require
a
KD4
‐
1/2
‐
A
coupler
socket.
Activate
the
compressed
‐
air
supply.
Step
5:
Place
a
container
underneath
the
valve,
to
collect
any
liquid
dripping
out.
The
container
needs
sufficient
capacity
to
contain
the
purifying
agent
together
with
the
solved
residual
liquid.
Step
6:
Initiate
the
purging
procedure
by
pressing
[F1]
.
Keep
the
[F1]
key
pressed
until
the
procedure
is
complete
and
no
more
liquid
flows
out
through
the
nozzle
unit.
INFORMATION
(automatic
closing)
After
ca.
2
min
the
valve
closes
automatically,
in
order
to
protect
the
actuator.
Alternatively,
the
procedure
can
be
performed
via
the
RS
‐
232C
interface.
Use
the
commands
VALVE:UP
and
VALVE:DOWN
for
this
purpose.
Step
7:
Separate
the
compressed
‐
air
connection
and
remove
the
cartridge.
Step
8:
Make
sure
the
collected
liquid
in
the
container
is
disposed
of
according
to
local
regulation.
Содержание MDS 3000 Series
Страница 1: ......
Страница 58: ...VTK VS BA 004e B User Manual MDS 3200A Cleaning 58 8 3 4 Fine Purification of Components...
Страница 71: ...VTK VS BA 004e B User Manual MDS 3200A Attachment 71 13 ATTACHMENT 13 1 EC Declaration of Conformity...
Страница 72: ...VTK VS BA 004e B User Manual MDS 3200A Attachment 72 13 2 Dimensional Drawing MDC 3200A...
Страница 73: ...VTK VS BA 004e B User Manual MDS 3200A Attachment 73 13 3 Dimensional Drawing MDV 3200A with MDX 3080 CC...
Страница 74: ...VTK VS BA 004e B User Manual MDS 3200A Attachment 74 13 4 Connection Diagram PLC Interface...