10.
TECHNICAL SPECIFICATION
Feedwater
Source
Potable tap water as detailed below.
Failure to comply with the minimum feedwater pretreatment recommendations
will affect the life and performance of key components within the
MEDICA
, and
may invalidate the warranty.
Contaminant
Measure
Range
Pretreatment
Calcium
Ca ppm as CaCO
3
< 250
None
> 250
Softener or
use very low RO recovery
Total chlorine
Cl ppm
< 0.1
Protek L1
0.1 to 0.5
Protek L1
> 0.5
Protek L2
Silica
SiO
2
ppm
< 30
Protek L1 or L 2
> 30
20 inch cartridge depth filter*
Fouling Index
FI
<20
Protek L1 or L2
> 20
Backwashable media filter with a
minimum flow rate of 20l/min
Iron/manganese
Fe/Mn ppm
< 0.1
Protek L1 or L2
> 0.1
Back-washable pre- filter *
Organics
TOC ppm C
<3
Protek L1 or L2
> 3
Cylinder of carbon sized correctly
TEMPERATURE
4 - 40ºC (Recommended 15 - 25ºC)
FLOWRATE
(maximum requirement @15
O
C)
9l/min
Drain requirements (gravity fall with air gap).
9l/min
Feedwater Pressure
6 bar (90 psi) maximum, 2 bar (30 psi) minimum.
*
Installed in feed water supply.
Note: If feedwater purity is variable or values are
close to the top of one of the ranges, provide
pre-treatment for the higher range or seek
advice from Technical Support at ELGA
LabWater.
Dimensions
Height
820 mm (32.8”) 834 mm (33.4”) including castors
Width
794 mm (31.8”)
Depth
470 mm (18.8”)
B
S
R
RE
Supply weight
44 kg
45 kg
52 kg
53 kg
Operational weight
101 kg
102 kg
109 kg
110 kg