ELGA
MEDICA Pro 30/60/120 US Operator Manual
`
MEDICA Pro 30/60/120 US MANU40594 Version 1
– 07/14
Page 17
4.
OPERATION
4.1
Night service/Operational day (screen 030)
Refer to Section 3.2 - Step 14 and 15.
Your
MEDICA
can be programmed to operate on specific days
between selected times. This is to optimize the efficiency of the unit
and to minimize rises in water temperature.
During the ‘sleep’ period the unit will display
.
It is possible to override this mode by PRESSING
.
During the ‘sleep’ period the unit will run in intermittent recirculation
(10 minutes every two hours) to maintain water purity around the
distribution loop.
4.2
Continuous Recirculation (24/7) (screen 029)
Refer to Section 3.2 - Step 13.
If the unit is set to continuous recirculation, it will constantly re-
circulate the water and fill the reservoir as required.
It is recommended that the system only runs in continuous mode
when the demand for water is high (greater than 50% of the make up
flow)
During recirculation the water temperature will tend to increase
slowly.