MEDICA Pro 30/60/120 US Operator Manual
ELGA
Page 26
MEDICA Pro 30/60/120 US MANU40594 Version 1 - 07/14
7.
EMERGENCY BYPASS
Contact your local service provider before
putting the unit into emergency bypass.
7.1
Emergency bypass (MEDICA R/RE/RX)
1.
Ensure the process is OFF
2.
Disconnect the electrical supply
3.
Open the doors.
4.
Close valve 1
5.
Close valve 2
6.
Remove yellow plastic cover 3
7.
Open valve 3
8.
Close valve 3 when there is no requirement for
purified water.
9.
Monitor water purity independently or install new
Medpure L1 cartridges to guarantee water purity.
7.2
Emergency bypass (MEDICA B/S/SE)
1.
Ensure the process is OFF
2.
Disconnect the electrical supply
3.
Open the doors.
4.
Close valve 1
5.
Valve 2 should remain closed
6.
Remove yellow plastic cover 3
7.
Open valve 3
8.
Monitor the level in the reservoir to ensure it does not
overflow.
9.
Monitor water purity independently or install new
Medpure L1 cartridges to guarantee water purity.