Connections
Inlet
10 mm OD tube
Drain
10 mm OD tube
Reservoir drain
10 mm OD tube
Recirculation loop outlet*
10 mm OD tube
Recirculation loop inlet*
10 mm OD tube
Electrical Requirements
Mains input
230 V ac, 50 Hz
115 V ac, 60 Hz
System control voltage (not including pumps and UV)
24 V dc
Power consumption (peak demand)
650 VA
Electrical protection rating
10 amp
Safety Features
Power fail safe
Water temperature alarm*
Water purity alarm*
Leak detection alarm
Access restricted by User Recognition key
Low voltage control circuit - 24 V dc
Visual alarms
Audible alarms
*
Output to operate 24 V dc relay.
Product Water Specification
Flowrate
MEDICA Pro 30
MEDICA Pro 60
MEDICA Pro 120
Daily usage - typical
240
480
960
Daily usage - maximum
720
1440
2880
Inorganic
>10 MΩ. cm @ 25
O
C
TOC ppb
<30
Bacteria
<1 CFU/ml**
Particles
0.05 µm
**
System to be regularly sanitized and installed following ELGA
LabWater installation design guidelines.
As part of our policy of continual improvement we reserve the right to
alter the specifications given in this document.