SWITCH on power. FOLLOW screen prompts to
accept new cartridge replacement date
9.
START unit and allow to circulate until water quality is
achieved.
CAUTION! If operating the unit with two Medpure L1
purification packs installed, both packs
must be replaced at the same time. Failure
to do so will result in low water purity.
5.3
Replacing Ultra-Microfilter (LC109)
The Ultra-Microfilter (UMF) should be replaced in the
following circumstances:
When indicated by the consumable alarm
After a maximum of 12 months use
Product flow is no longer adequate
Product water bacterial content is too high.
1. ISOLATE power
2. OPEN front doors and LOCATE UMF
3. UNSCREW lower connection
Note: At this stage a small quantity of water may spill.
4. UNSCREW UMF from top connection
5. INSTALL new UMF
6. RECONNECT
7. RINSE water to drain to recommission the system.
8. RESET reminder as described in section 3.3