MEDICA Pro 30/60/120 US Operator Manual
ELGA
Page 4
MEDICA Pro 30/60/120 US MANU40594 Version 1 - 07/14
1.6
Sanitization Chemicals
During the sanitization cycle Minncare Cold Sterilant is used and
relevant safety guidance is included in this handbook. Please refer to
the manufacturer for material safety data sheets.
Minncare Cold Sterilant is EPA registered as a sterilant, high level
disinfectant, and sanitizer.
Spent Minncare Cold Sterilant is acidic and requires normal
neutralization as specified by your local state and local regulations.
1% of Minncare Cold Sterilant has a pH of 3.5.
1.7
Control of Substances Hazardous to Health
(COSHH)
Material safety data sheets covering the various replaceable
components are available upon request. Contact your local ELGA
LabWater distributor.
1.8
Consumables.
Consumables
Typical
Service Life*
Max. Shelf
Life
LC105 (UV lamp)
12 months
5 years
LC109 (UMF)
12 months
2 years
LC136 (Vent filter)
6 months
2 years
LC174 (Medpure L1 cartridge)
3-12 months
2 years
LC175 (Protek L1)*
6 months
2 years
LC 177 (Protek L2)*
12 months
2 years
LC180 (RO cartridge 60 L/hr)
3 years
2 years
LC181 (E-Cartridge)**
3 years
2 years
* Service Life is an estimate only, and will depend on the
application and feed water quality. Care should be taken to order
the correct consumable items.
** Optional - Consumables not fitted as standard. Fit only after
consultation with ELGA Technical Support