MEDICA Pro 30/60/120 US Operator Manual
ELGA
Page 18
MEDICA Pro 30/60/120 US MANU40594 Version 1 - 07/14
5.
MAINTENANCE
An approved supplier or distributor should carry out any
maintenance work not included in this han
d
book.
Note: Disposal of all end of life consumable items should
be in accordance with local statutory regulations
WARNING! ALWAYS
CHECK
THAT
THE
MAINS
ELECTRICAL POWER AND FEED WATER
ARE SWITCHED OFF BEFORE ATTEMPTING
ANY MAINTENANCE PROCEDURE.
5.1
Replacing Composite Vent Filter (L136)
The Composite Vent Filter (CVF) should be replaced in the
following circumstances:
When indicated by the alarm (screen prompt) or after a
maximum of six months.
1.
ENSURE process is OFF and isolate power.
2.
OPEN front doors and locate CVF.
3.
UNSCREW old CVF and discard
4.
UNPACK new CVF.
5.
WRITE the installation date on the label of the filter for
future reference.
6.
INSTALL filter.
7.
SWITCH on power.
8.
RESET consumable reminder as described in Section
3.3.
5.2
Replacing Medpure L1 (LC174)
When the purity of water from the unit starts to
deteriorate.
When indicated by the consumable alarm or after
a maximum of six months.
1.
ENSURE process is OFF and isolate power.
2.
OPEN right hand door and LOCATE Medpure
cartridge.
3.
PUSH Medpure FORWARD, then LIFT, finally PULL
to remove the used cartridge. Discard.
4.
UNPACK new Medpure.
5.
REMOVE sealing plugs from inlet and outlet ports.
6.
Wet ‘O’ rings and SLIDE new cartridge into position
pushing upwards against pack reader contacts.
7.
EASE back and ENSURE that the pack is fully
engaged (down) in the retainers.