MEDICA Pro 30/60/120 US Operator Manual
`
MEDICA Pro 30/60/120 US MANU40594 Version 1
– 07/14
Page 11
Step 12 - Product water temperature alarm settings TS2
(screen 028)
This screen is used to select the value at which the product
water temperature alarm will activate. The alarm does not
stop the unit and will automatically reset if the temperature
returns below the set point.
1.
PRESS
to select alarm point (increments of 1,
ranging from 20 to 50
C)
OR
PRESS
✔
to proceed to step 13.
2.
PRESS
✔
.
Note: If the water temperature rises above 55
C in the
system it will alarm, stop and await operator
intervention - alarm will be reset once power is
removed for 5 seconds and then reinstated
Step 13 - Continuous operation (screen 029)
Continuous operation may be required in particular
circumstances or when demand for water fluctuates.
CAUTION! During long periods of continuous operation
and low usage the water temperature will
rise. It is recommended that this function is
only used when water usage is on average
>50 litres/hr and water is used every day.
1.
PRESS
to change (
= ON)
OR
PRESS
✔
to proceed to step 14.
Note: Proceed to step 16 if continuous operation is
selected (
= ON).
Step 14 - Periods of operation (screen 030)
For efficiency and to reduce the likelihood of heat build up
the normal operational hours can be selected.
During ‘OFF’ periods the unit will automatically re-circulate
water for a period of 10 minutes every two hours this will
maintain water purity within the distribution loop.
1.
PRESS
to edit night service start
OR
PRESS
✔
to proceed to step 15.
2. PRESS
to increase or
to decrease time in
increments of 30 minutes.
3.
PRESS
to step to night service end.
Setting period of operation
Continuous operation screen
Alarm settings TS2 screen