MEDICA Pro 30/60/120 US Operator Manual
ELGA
Page 22
MEDICA Pro 30/60/120 US MANU40594 Version 1 - 07/14
5.6
Replacing E- Cartridge (LC181)
standard on RE
variant only
The E- cartridge should be replaced:
When indicated by the alarm
1. ENSURE process is OFF and ISOLATE power.
2. OPEN front doors
3. LOCATE E
– cartridge
4. DISCONNECT tubing to air pump and inlet and outlet
tubing from the E cartridge.
5. REMOVE used cartridge from clip and discard in
accordance with local regulations.
6. TAKE new cartridge
7. REMOVE sealing plugs from inlet and outlet ports
8. LOCATE new cartridge in clip
9. RE-CONNECT tubing
5.7
RO modules
The reverse osmosis modules should be replaced if
permeate water purity or flow rate is not adequate and does
not meet predicted or previous performance. For information
regarding the replacement of the reverse osmosis module
contact Customer Services.
E-Cartridge position
E- Cartridge
All manuals and user guides at all-guides.com