4
SETTINGS AND CALIBRATIONS
37
Chapter
4
SETTINGS AND CALIBRATIONS
6.
Increment
Set the minimum unit of load display. Loads are displayed in the
set increments.
Operation
Setting range (1, 2, 5, 10)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Next
→
Increment
7.
Analog Filter
The low-pass filter of the set cut-off frequency is inserted with
respect to the input of the torque meter.
Operation
Setting range (10, 30, 100, 300 Hz)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Next
→
Next
→
Analog Filter
8.
Digital LPF
The low-pass filter of the set cut-off frequency is digitally
processed and inserted with respect to the input of the torque
meter.
Operation
Setting range (0(OFF) to 30 Hz)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Next
→
Next
→
Digital LPF
9.
Digital Filter
Loads are moving-averaged by the set number of times.
Operation
Setting range (0, 2 to 999 Times)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Next
→
Next
→
Digital Filter
10.
Digital Zero Limit
Set the limit of load that can be digital-zeroed. If Digital Zero is
performed with a load larger (smaller) than the zero-calibrated
load by the set load, an error will result, and the load will not be
zeroed.
Operation
Setting range (0 to 9999)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Next
→
Next
→
DZ Limit
Содержание TM500
Страница 1: ...01JUN2016REV 1 02 TORQUE MONITOR TM500 OPERATION MANUAL ...
Страница 9: ...Contents VIII M E M O ...
Страница 133: ...5 COMMUNICATIONS 124 Chapter 5 COMMUNICATIONS About messages Unless otherwise noted all the messages are in ASCII code ...
Страница 149: ...8 SPECIFICATIONS 140 Chapter 8 SPECIFICATIONS 8 2 Outside dimensions TM500 TORQUE MONITOR ࠉ ࠉ ࠉ Unit mm ...
Страница 165: ......