4
SETTINGS AND CALIBRATIONS
65
Chapter
4
SETTINGS AND CALIBRATIONS
⑤
To input a waveform again, repeat
②
to
④
.
⑥
Pressing
registers the reference waveform(s).
In the case of one reference waveform, it is
recognized that the HI limit waveform and LO limit
waveform are the same.
If two or more waveforms are input, the upper limit
of all waveforms becomes the HI limit waveform,
and the lower limit of all waveforms becomes the
LO limit waveform.
2.
Compare Area
Set the waveform comparison target.
Setting range (start point, end point)
Time or displacement (Front)
(0 to 2047 × Measurement Length/2000)
Displacement (Back)
Operation
(-2047 to 0 × Measurement Length/2000)
Main screen
→
Setting
→
Work Setting
→
Wave Setting
→
Compare Area
①
To change the start point of the area, press
(Displacement (Back):
), and
then move the cursor to specify it.
②
To change the end point of the area, press
(Displacement (Back):
), and
then move the cursor to specify it.
③
To register the specified area, press
.
④
After the registration of the area, the background color outside the area becomes gray on the
waveform processing screens set later so that the boundaries of the area become clear.
Load
1st waveform
Time or displacement
Load
1st waveform
Time or displacement
2nd waveform
3rd waveform
Содержание TM500
Страница 1: ...01JUN2016REV 1 02 TORQUE MONITOR TM500 OPERATION MANUAL ...
Страница 9: ...Contents VIII M E M O ...
Страница 133: ...5 COMMUNICATIONS 124 Chapter 5 COMMUNICATIONS About messages Unless otherwise noted all the messages are in ASCII code ...
Страница 149: ...8 SPECIFICATIONS 140 Chapter 8 SPECIFICATIONS 8 2 Outside dimensions TM500 TORQUE MONITOR ࠉ ࠉ ࠉ Unit mm ...
Страница 165: ......