4
SETTINGS AND CALIBRATIONS
36
Chapter
4
SETTINGS
AND CALIBRATIONS
1.
Unit
Select a unit.
Operation
Setting range (See "9-3.Unit setting list"on p.141.)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Unit
2.
Zero Calibration
Register the present sensor input as zero.
Press
with the torque meter unloaded.
Operation
Setting range (-5.555 to 5.555 V)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Zero Cal.
3.
Equivalent Input Calibration
Input the rated output and rated capacity (display value)
described on the data sheet of the torque meter.
Also, set the decimal place here. Input the decimal point
together when inputting the display value.
Setting range
Rated output
(-9.999 to 9.999 V)
Operation
Display value (rated capacity) (-9999 to 9999)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Next
→
Equiv. Cal.
4.
Actual Load Calibration
Apply actual load to the torque meter, and input the load value at
that time.
Also, set the decimal place here. Input the decimal point
together when inputting the calibration value.
Setting range
Calibration value
(-9999 to 9999)
Operation
Display value
(display only)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Next
→
Actual Cal.
5.
Overload
Set the load at which the sensor is judged to be overloaded. The
Load OK output turns OFF when the load exceeds the set value.
Operation
Setting range (0 to 9999)
Main screen
→
Setting
→
First Setting
→
Y-axis Setting
→
Next
→
Overload
Содержание TM500
Страница 1: ...01JUN2016REV 1 02 TORQUE MONITOR TM500 OPERATION MANUAL ...
Страница 9: ...Contents VIII M E M O ...
Страница 133: ...5 COMMUNICATIONS 124 Chapter 5 COMMUNICATIONS About messages Unless otherwise noted all the messages are in ASCII code ...
Страница 149: ...8 SPECIFICATIONS 140 Chapter 8 SPECIFICATIONS 8 2 Outside dimensions TM500 TORQUE MONITOR ࠉ ࠉ ࠉ Unit mm ...
Страница 165: ......