- Continuous mode
When repeated twice at HI limit: three times and LO limit: twice
t1: Delay time between the instant when the Start input is turned ON and the instant when all outputs but
Run are turned OFF ... Approx. 5ms
t2: Delay time between the instant when the Start input is turned OFF and the instant when the measurable
section is started... 0.25ms
Displacement positioning is automatically executed, and measurement starts as a measurement start
point.
t3: Minimum time required for measurement completion at the next repetitive displacement... 0.5ms
t4: Delay time between the instant when the measurement is stopped and the instant when the Result
output is turned ON ... Approx. 50ms (However, there are differences between measurements.)
t5: Delay time between the instant when the Result output is turned ON and the instant when the Complete
output is turned ON ... Approx. 30ms
- Turning OFF
→
ON the external input signal “Start” releases the hold, and turns OFF all outputs
but Run. (Reset operation by the Start input)
- After turning “Measurement completion” from OFF to ON, “Complete” turns ON if either
measured until repetitive displacement or measurement completion condition is satisfied.
- Once the judgment of output becomes LO or HI, the output is maintained.
- OK turns ON if the Compare Area is passed through during measurement and the result is not LO
or HI after completion of measurement.
- The point where HI or LO was detected at the last acquired waveform is a judgment point.
Load
Compare area
Wave result
Stop input
OFF
ON
Start input
OFF
ON
Complete output OFF
ON
OK output
Wave result
LO or HI output
Measurement section
Calculating
Output OFF
HI output
Output OFF
OK is output if
result is not LO or HI.
Reset operation
t4
by the Start input
t3
t1
t2
Waveform
comparison
Displacement
Repetitive displacement
Compare area
Measurement section
Repetitive displacement
HI/LO output
Displacement
OFF
ON
positioning
t5
1
1
2
2
3
4
1
2
3
1
2
Sensor
input value
Result of the
previous
measurement
Result of the
previous
measurement
judgment point
(automatic execution)
Содержание TM500
Страница 1: ...01JUN2016REV 1 02 TORQUE MONITOR TM500 OPERATION MANUAL ...
Страница 9: ...Contents VIII M E M O ...
Страница 133: ...5 COMMUNICATIONS 124 Chapter 5 COMMUNICATIONS About messages Unless otherwise noted all the messages are in ASCII code ...
Страница 149: ...8 SPECIFICATIONS 140 Chapter 8 SPECIFICATIONS 8 2 Outside dimensions TM500 TORQUE MONITOR ࠉ ࠉ ࠉ Unit mm ...
Страница 165: ......