2.CONNECTION
8
2-5. Power Input Terminal Connection
Connect an AC input cord. The input power source is free in the range of 100 to 240V AC (-15%,
+10%). The frequency is 50/60Hz.
Make connections to the terminal block with a solderless terminal within 6mm in diameter as shown in
the illustration so as not to let the tip of the cable spread out.
1. Make sure that no power is applied.
2. Remove the screws (two), and remove the terminal block
cover.
3. Remove the two screws(“L” and “N”) at the terminal block.
4. Align the solderless terminals with the screw holes, and then
tighten the screws.
5. Install the terminal block cover, and fix it with the screws
(two).
2-6. Protective Ground Connection
The grounding terminal is for prevention of electric shocks.
Use an approx. 0.75mm
2
thick wire, and be sure to ground.
1. Make sure that no power is applied.
2. Remove the screws(M4) at the protective ground.
3. Align the solderless terminals(M4) with the screw holes, and
then tighten the screws.
Within 6mm
WARNING
- Connect with no power applied because it may cause an electric shock.
- The attached AC cable is designed for domestic use in Japan, and its rating is 125V
AC, 7A. For use at voltages exceeding the rating and for overseas use, have a
separate AC cable prepared.
- Since the F600AT has no power switch, install a breaker.
- Be sure to ground the protective ground terminal to prevent electric shocks.
Do not use other screws than that attached to the main body.
Protective ground
M4