4.CALIBRATION

19

4-3-6. Span calibration

Set the actual load value under an actual load.

Setting range:

00001 to 99999 (The decimal point is selectable.)

4-4. Net Over/ Gross Over

When the net weight value or gross weight value exceeds a fixed regulation, an alarm is given by this
function.

NET OVER:

00000 to 99999

GROSS OVER:

00000 to 99999

The alarm is given when the following conditions are met.

4-5. Unit Setting

Set the unit of the load to be displayed.
The display value (calibration value) is not affected by changing the unit.

Alternatives:

NONE

 

kg

 

t

 

g

 

N

 

lb

4-6. Tare Subtraction

The net weight is zeroed by this function. The gross weight value is not
changed by this operation.

Alternatives:

RESET

 

NO

 

YES

: Perform the tare subtraction.

Go back to the previous screen without executing tare subtraction.

Reset the tare subtraction

.

MODE

CALIBRATION

SPAN CAL.

Once

MODE

 Once

CALIBRATION

NET OVER

Once

MODE

 Once

CALIBRATION

GROSS OVER

Once

Conditional expression

Display

NET OVER

Net weight 

 “NET OVER” set value

NET OVER

GROSS OVER

Gross weight 

 “GROSS OVER” set value

GROSS OVER

MODE

Once

CALIBRATION

UNIT DISPLAY

Once

TARE

  

ON

* It operates it with the net weight (NET) displayed. 

YES

NO

RESET

Содержание F600AT

Страница 1: ...F600AT 01MAY2015REV 3 01 WEIGHING CONTROLLER OPERATION MANUAL...

Страница 2: ...rmed by personnel having technical knowledge of electricity In order to have an F600AT used safely notes I would like you to surely follow divide into and and are indicated by the following documents...

Страница 3: ...The attached AC cable is designed for domestic use in Japan and its rating is 125V AC 7A For use at voltages exceeding the rating and for overseas use have a separate AC cable prepared Before perform...

Страница 4: ...led Warning during startup and maintenance CAUTION Use the F600AT as it is incorporated in a control panel etc Do not install in the following environments Locations where temperature or humidity exce...

Страница 5: ...at least 30 minutes or more before use If the F600AT is not used by the specified method its protective performance may be impaired Maintenance When performing maintenance disconnect the power Do not...

Страница 6: ...Gauge Sensor Cable Coloration 7 2 5 Power Input Terminal Connection 8 2 6 Protective Ground Connection 8 2 7 SI F Connection 9 2 8 External I O Connection 9 2 8 1 How to assemble the connector 9 2 8...

Страница 7: ...ilter Motion detect period Motion detect range Display frequency Zero tracking period Zero tracking range Indicate color Backlight lighting time Backlight bright and dark time Display table selection...

Страница 8: ...BCD DATA OUTPUT OPTION 51 15 1 Connector Pin Assignment 51 15 2 Equivalent Circuit 52 15 3 Signal Input 52 15 4 Signal Output Timing 53 15 5 BCD Data Output Setting 54 Data output rate Output weight...

Страница 9: ...CONTENTS VIII 21 SETTING ITEM LIST 65 22 SPECIFICATIONS 70 Analog Section Display Section Setting section Communication Option External Input Output General Performance...

Страница 10: ...OR DIGITAL INDICATOR DIGITAL INDICATOR DIGITAL INDICATOR DIGITAL INDICATOR DIGITAL INDICATOR DIGITAL INDICATOR DIGITAL INDICATOR DIGITAL INDICATOR DIGITAL INDICATOR WEIGHING CONTROLLER 15JAN2008REV 1...

Страница 11: ...1 3 2 Rear Panel Weight display Graph display Lamp indication Status Green lighting Steady state Orange lighting Writing into internal NOVRAM Do not turn off the power of the F600AT Yellow blink Volt...

Страница 12: ...ble The power cable attached to this product as standard equipment can be used in the AC100V power supply in Japan Official ratings voltage AC125V Please use the power cable authorized in the country...

Страница 13: ...vent electric shocks RS 232C connector RS 232C connector to transmit receive measurement data and status information etc The compatible connector is HOSIDEN manufactured TCP8080 015267 or equivalent O...

Страница 14: ...iver 3 to 3 5mm 1 in shaft diameter is recommended Precision screwdriver etc 4 Insert the wire into the connection hole so as not to let the tip spread out 5 Tighten the screw with the screwdriver 0 5...

Страница 15: ...in excitation voltage over long cable runs 4 wire connection Connect 3 and 4 and 5 and 6 as shown below Be sure to connect 4 and 6 on the terminal block with the accessory jumper lines because if the...

Страница 16: ...d check signal names and colors in order to connect the cables correctly R R R R R R EXC SIG SIG EXC EXC SIG EXC SIG SHIELD S S n number loadcells connected in parallel are considered one unit whose c...

Страница 17: ...the screws two 2 6 Protective Ground Connection The grounding terminal is for prevention of electric shocks Use an approx 0 75mm2 thick wire and be sure to ground 1 Make sure that no power is applied...

Страница 18: ...distance is approximately 300m Do not parallel it with AC lines and high voltage lines It may cause of malfunction Please refer to 13 AUTOMATIC PRINT on page 45 for print 2 8 External I O Connection...

Страница 19: ...ed by means of a contact such as a relay or a switch or a noncontact such as a transistor or an open collector TTL COM Spark Killer Load F600AT Inside Spark Killer DC Varistor Load AC Relay Vext Suppl...

Страница 20: ...e 1 2 3 4 5 6 7 8 Case DCD DSR RXD TXD DTR GND RTS CTS 1 6 2 3 4 5 7 8 4 3 2 1 6 5 8 7 9 DTR TXD RXD DCD DSR GND CTS RTS RI TCP8080 015267 or equivalent D Sub 9pin F600AT PC etc Cabling diagram CA81 2...

Страница 21: ...Configuration O K ESC ESC ESC BACK BACK BACK BACK or or MODE ESC O K Set value entry from the ordinary display screen Ordinary display screen Mode setting screen EACH CODE Mode Keys Set value selectio...

Страница 22: ...on 1 Press the key on the ordinary display screen 2 The mode setting screen appears Select the mode Modes are as follows EACH CODE COMPARISON OPERATION GRAPH OPTION SYSTEM CALIBRATION SEQUENCE MODE RS...

Страница 23: ...9 INDICATE COLOR P 29 BACK LIGHT P 30 PAGE3 HI LO COMP SEL P 25 NEAR ZERO COMP P 25 PRESET TARE WGT 1 P 25 NEAR ZERO P 25 PRESET TARE WGT 2 P 25 PAGE3 DISPLAY SELECT P 30 HI LO OUT SEL P 31 LOCK1 P 31...

Страница 24: ...CK DSP1 P 60 DELFCHECK MEM P 60 SELFCHECK KEY P 60 SELFCHECK EXT P 60 PAGE2 SELFCHECK DSP2 P 60 SELFCHECK COM P 60 PASSWORD P 34 LANGUAGE P 34 GROSS NET KEY P 34 SYSTEM PAGE1 COMM MODE P 46 BAUD RATE...

Страница 25: ...ated of the strain gage sensor by the keys Calibration is easily performed when no actual load is available For example the gain is automatically determined by entering as indicated for a load Actual...

Страница 26: ...alibration Procedure Follow the steps below to perform equivalent input calibration Release the calibration protection Set the unit of the load to be displayed Determine the decimal point place of the...

Страница 27: ...m the following Alternatives NONE 0 0 0 00 0 000 4 3 4 Zero calibration Set the zero point in no load condition 4 3 5 Equivalent input calibration Set the rated output value and reading of the sensor...

Страница 28: ...is not affected by changing the unit Alternatives NONE kg t g N lb 4 6 Tare Subtraction The net weight is zeroed by this function The gross weight value is not changed by this operation Alternatives R...

Страница 29: ...lue the alarm ZALM is displayed when operation is performed with any gross weight value exceeding the range Setting range 0000 to 9999 The displayed ZALM can be invalidated by the following method DZ...

Страница 30: ...rate discharge of the weight value Comparisons are made with the net weight High low limit comparisons are made with the net or gross weight Comparison inhibiting time Judging time Completion output t...

Страница 31: ...cient C and the value is added to or subtracted from the preset fall value In order to minimize the errors the values of D D1 D2 D3 DA can also be regulated Setting of automatic fall parameters AFFC S...

Страница 32: ...00 0 000 2 0 530 7 20 010 0 010 3 0 530 8 20 110 0 110 3 0 530 9 20 010 0 010 4 0 0 530 0 040 4 0 010 0 010 2 4 0 005 CPS Value 10 19 880 0 120 1 0 535 11 19 990 0 010 1 0 535 12 20 010 0 010 2 0 535...

Страница 33: ...f the completion output time after the weight value has become stable and the judging time has elapsed Overweight underweight comparison mode Valid when DIS O U is selected by COMP MODE SEL Select the...

Страница 34: ...play NZ lights and the NZ output turns ON By selecting COMP OFF the NZ output is turned OFF NEAR ZERO COMP GROSS NET COMP OFF ABS GROSS ABS NET NEAR ZERO 00000 to 99999 Tare setting selection Preset t...

Страница 35: ...y comp mode sel When the weight value HI HI limit the status display HH lights and the signal output HH SP1 turns ON When the weight value HI limit the status display HI lights and the signal output H...

Страница 36: ...ating function 2 Output in the simple comparison mode For the sequence mode refer to page 40 For final discharge HI LO limit comparison when DIS H L is selected by comp mode sel When the net final wei...

Страница 37: ...natives 2Hz 4Hz 6Hz 8Hz Motion detect period Motion detect range Set the parameter to detect stability When the width of change in weight value falls below the preset range minimum scale division and...

Страница 38: ...tracking automatically takes the zero point at the preset intervals when the amount of movement of the zero point is less than the preset range minimum scale division The period tracking delay can be...

Страница 39: ...hree combinations The combinations vary with comparison modes Alternatives F SP1 3 F SP3 H L F SP3 T Z When DIS H L is selected Explanation of combination in each comparison mode For HI LO limit compa...

Страница 40: ...imit weight value GO ON when HH HI LO and LL are all OFF LO ON when the weight value preset low limit LL ON when the weight value preset LO LO limit Signal name Condition HH ON when the preset HI HI l...

Страница 41: ...ber of drawing points Y axis Y axis is load axis setting Drawing is performed from the value set by the Y axis start point to the value set by the Y axis end point The number of drawing points is 100...

Страница 42: ...me end point Set the time to display one screen The setting range is from 2 sec to 98 sec at intervals of 2 sec The X axis time end point can also be set from the graph screen Setting range 2 to 98 se...

Страница 43: ...ntenance and inspection Do not operate Language The display language is selectable between Japanese and English JPN JAPANESE ENG ENGLISH GROSS NET key Set OFF ON of the key operation of key Alternativ...

Страница 44: ...e B4 Function Selection setting makes the following effective KEY_LOCK When the terminal is short circuited all key operations are disabled When KEY_LOCK terminal function is selected measurement code...

Страница 45: ...ode ADJUST FEEDING OFF ON ADJ FEEDING TIME 0 00 to 9 99 sec What is Adjust feeding If the Net Final weight Under weight at the completion signal ON timing when the judging time has elapsed after the S...

Страница 46: ...times When the start signal is input if 1 the net final weight SP1 or 2 the final weight CPS 0 the feeding gate signal SP1 SP2 or SP3 is not turned ON but a sequence error results Sequence error 5 1 i...

Страница 47: ...lue SP3 output ON when the net preset final weight value preset CPS value UNDER output ON when the net preset final weight value preset underweight value OVER output ON when the net preset final weigh...

Страница 48: ...ght value OVER output ON when the net preset final weight value preset overweight value Select the weight value for near zero comparison from GROSS or NET under COMPARISON setting menu LO limit NET 0...

Страница 49: ...lue SP3 output OFF when the net weight preset final weight value preset CPS value UNDER output ON when the net weight preset final weight value preset underweight value OVER output ON when the net wei...

Страница 50: ...ut it is output at the OFF edge ON OFF of the final weight signal SP3 signal Completion output time Adjust feeding time Judging time Comparison inhibiting time Judgment output END output SP1 output ON...

Страница 51: ...raction is performed at the intervals of these times When the number of AZ times is set at 00 the AZ function is turned OFF When the number of judging times is set at 01 judgment is made at each time...

Страница 52: ...ON Tare subtraction Edge input B7 The ON edge OFF ON of the external input will result in immediate tare subtraction to zero the net weight 5 TARE OFF Tare subtraction reset Edge input B8 The ON edge...

Страница 53: ...SP3 Comparison output A6 6 END Completion A7 This indicates measurement completion output which is valid only in the carving out mode 7 NZ Near zero A8 This is near zero comparison output Refer to pa...

Страница 54: ...he near zero signal is OFF If the automatic print is output it can next be output after the near zero signal is once turned ON 13 2 In The Discharge Mode Simple Comparison Mode The command is output a...

Страница 55: ...or 2 bit selectable Parity None odd or even selectable Terminator CR CR LF selectable Code ASCII 14 2 RS 232C Interface Setting Set the RS 232C communication conditions of the F600AT Communication mo...

Страница 56: ...it Host F600AT R A R A 1 0 0 0 0 Terminator CR Host F600AT R B R B 1 0 0 0 0 Terminator CR Host F600AT R C R C 1 0 0 0 0 Terminator CR NET display GROSS display In tare subtraction Not in tare subtrac...

Страница 57: ...ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF 1 ON 0 OFF Sequence error 0 to 5 LOAD occurred LOAD occurred GROSS OVER occurred NET OVER occurred No error Host F600AT R G R G 0 0 0 0 0 0 0 Term...

Страница 58: ...equivalent input calibration and weight value Therefore check the rated output and rated capacity of the load cell as in the case of equivalent input calibration by the touch panel Example It is assu...

Страница 59: ...smitted 1 H In holding O Weight error S Stable M Unstable 2 T Zero tracking ON A Zero tracking OFF 3 For HI LO limit comparison For final weight DIS O U and final weight DIS H L F Comparison OFF G OVE...

Страница 60: ...tible connector FCN 361J032 AU manufactured by FUJITSU COMPONENT or equivalent Connector cover FCN 360C032 B manufactured by FUJITSU COMPONENT or equivalent Connector cover optional type CN51 Special...

Страница 61: ...n shorted with COM No strobe pulse is output F600AT Inside Vcc 5V COM Vext Vceo 30V max Ic 50mA max Internal transistor status Output data Negative Positive 0 OFF ON 1 ON OFF Output pin level Output d...

Страница 62: ...trol mode sequence mode The signal is output at the ON edge of the completion signal at the intervals of judging times OVER The signal is output with ADC OVER LOAD or ADC OVER LOAD overflow NET OVER o...

Страница 63: ...ut weight Alternatives INDICATE GROSS NET Select the weight value for BCD output If INDICATE is selected the weight indicated by the indicator is output 15 6 Self Check The BCD input output status can...

Страница 64: ...00 times sec The output has an overrange of approx 10 FS Either voltage output or current output can be mounted Taking voltage output signals Connect external equipment 2k or more load resistance to a...

Страница 65: ...e to fit the terminal hole 2 Insert the supplied screwdriver into the upper hole and lift upward 3 Insert the twisted wires into the lower hole 4 Pull the screwdriver out from the upper hole 5 Make su...

Страница 66: ...e should be smaller than the preset full scale value How to adjust Adjust the output with the dummy trimmer that appears on the adjustment screen after selecting each fixed output in the D A output mo...

Страница 67: ...t point is larger than the set value of the Y axis load end point Remedy Set again so that the Y axis load start point becomes smaller than the Y axis load end point LOCK IS ON Cause Key operation is...

Страница 68: ...early 3 Discharged items are jammed SEQ ERR 5 Cause 1 The SP1 output is ON when weighing is started On condition that it is set to confirm the weight value when weighing is started 2 Final weight CPS...

Страница 69: ...ched square turns yellow Pressing the key at the upper right of the screen restores the previous screen Self check EXT Check the operations of the external input output signals Each output turns on wh...

Страница 70: ...Port Serial Port Port Watch Dog CPU 32bit RAM 16kbyte NOV RAM 4096bit LCD Driver LCD Flash ROM SI F DeviceNet BCD 256kbyte EXC EXC SIG SIG S S Analogue External SRAM External Flash ROM Serial Port RS...

Страница 71: ...20 DIMENSIONS 62 20 DIMENSIONS Standard equipment When the external input output signals is connected Unit mm Front Side Rear Top Unit mm Front Side Rear Top...

Страница 72: ...20 DIMENSIONS 63 When the BCD option is equipped When the CC Link option is equipped Unit mm Front Side Rear Top Unit mm Front Side Rear Top...

Страница 73: ...t dimensions 2 Remove the screws two and remove the guide rails from both sides Do not use other screws than those installed to the F600AT body 3 Fit in the F600AT from the front of the panel 4 Instal...

Страница 74: ...on mode DIS O U PAGE Parameter Initial value L1 L2 S N Alternatives Setting range Communication Command No CODE 0 0 7 00 R W 1 1 FINAL 000 00 00000 99999 11 R W 2 SP1 000 00 00000 99999 12 R W 3 SP2 0...

Страница 75: ...ZERO COMP GROSS 0 GROSS 1 NET 2 COMP OFF 3 ABS GROSS 4 ABS NET 3C R W 13 PRESET TARE WGT 1 ON 0 OFF 1 ON 3D R W 14 NEAR ZERO 000 00 00000 99999 3E R W 15 PRESET TARE WGT 2 000 00 00000 99999 3F R W OP...

Страница 76: ...ase of used D A OUTPUT PAGE Parameter Initial value L1 L2 S N Alternatives Setting range Communication Command No 1 1 D A OUTPUT MODE GROSS 0 GROSS 1 NET 2 ZERO SCALE 3 FULL SCALE 2 ZERO OUTPUT 000 00...

Страница 77: ...PFF KEY ON 0 OFF 1 ON 6D R W 14 START STOP KEY ON 0 OFF 1 ON 6E R W 15 B4 FUNC SELECT CODE2 6F R W CALIBRATION PAGE Parameter Initial value L1 L2 S N Alternatives Setting range Communication Command N...

Страница 78: ...G TIME 1 00sec 0 00 9 99 95 R W 2 6 AZ TIMES 01times 00 99 96 R W 7 JUDGING TIMES 01times 00 99 97 R W RS 232C PAGE Parameter Initial valueInitial value L1 L2 S N Alternatives Setting range Communicat...

Страница 79: ...igit Sign Minus sign display Minimum scale division Can be set in the range of 1 to 100 Decimal point 8 8 8 8 8 The display position can be set Over scale A D converter input over LOAD A D converter i...

Страница 80: ...hen four station is occupied DeviceNet Baud rate 125k 250k 500kbps Node address 0 to 63 External Input Output External output signal Transistor open collector output Emitter COM terminal 9 points The...

Страница 81: ......

Отзывы: