22.SPECIFICATIONS
70
22. SPECIFICATIONS
■
Analog Section
Sensor excitation
10V DC
Output current: Within 120mA
6-wire (Up to four 350
Ω
load cells can be connected.)
Signal input range
-0.2 to 3.0mV/V
Zero adjusting range
-0.2 to 3.0mV/V
Accuracy
Nonlinearity
Within 0.02%FS ±1digit (When 3mV/V is input)
Zero drift
Within 0.25
μ
V/
℃
RTI
Gain drift
Within 25ppm/
℃
A/Dconverter
Speed
100times/sec.
Resolution
24bit (binary)
Analog filter
Bessel type low-pass filter (-12dB/oct)
Cutoff frequency
2Hz, 4Hz, 6Hz, 8Hz (Selectable by setting)
Calibration
Equivalent input calibration and actual load calibration
■
Display Section
Display
3.5 inch TFT color LCD module (Display area: 71mm×53mm)
320×240 dots
Weight value display
5-digit
Sign: Minus sign display
Minimum scale division
Can be set in the range of 1 to 100.
Decimal point
8 8
.
8
.
8
.
8 The display position can be set.
Over scale
A/D converter input over:
“LOAD”
A/D converter input minus over :
“-LOAD”
Net weight exceeding the “NET OVER” setting:
“NET OVER”
Gross weight exceeding the “GROSS OVER” setting: “GROSS OVER”
Unit display
NONE, kg, t, g, N, lb is selectable.
Status display
Various states of weight are displayed.
■
Setting section
How to set
Set by touch panel operation (with a key click buzzer).
External setting can also be made by RS-232C etc..
Memory of set values
Initial values
NOV RAM (nonvolatile RAM)
Other values
Lithium-battery-backed-up C-MOS RAM
(The memory life is approx. 7 years or more,
which varies depending on the operating conditions
and storage environment.)
Protection of set values
Setting operation can be prohibited to prevent changes by misoperation.
(LOCK1, LOCK2)