10.BCD DATA OUTPUT
79
10-2. Sink Type (BCO Option)
Output
Output signal:
Indicated value data (5-digit), near zero, minus, over,
stable, strobe
Output logic:
Switching between positive logic and negative logic
Output type:
Sink type
When signal ON, output transistor ON.
When the input unit such as PLC is connected, plus common
is connected.
Rated voltage:
30V
Rated current:
50mA
Insulation method: Photo coupler insulation
Input
Input signal:
BCD data hold, logic switching
Input type:
Dead-front contact input (self-contained power supply)
Relay, switch, and transistor, etc. can be connected.
The signal is inputed by short-circuited and
opening the input terminal of the terminal COM.
When the transistor is connected, NPN output type (sink type)
is connected.
Current when short-circuited
About 6mA
Insulation method: Photo coupler insulation