4.SETTING MODE CONFIGURATION
27
Setting Mode4
*Decimal point position synchronizes with the setting of decimal place of setting mode 3.
Item
Default
Setting range
Set Value
LOCK
Calibration
LOCK
Reference
page
1
RS-485
Commu-
nication
Method
Baud rate
4
5 : 38400bps
4 : 19200bps
3 : 9600bps
2 : 4800bps
1 : 2400bps
0 : 1200bps
○
Character length
1
Fixed at 8 bits
Parity bit
2
2 : Even
1 : Odd
0 : None
Stop bit
0
Fixed at 1 bit
2
RS-485 ID
01
01 to 32
○
3
RS-485 Transmission
DeLay Time
00
00 to 99
○
4
BCD Data Update Rate
4
4 : 2000 times
3 : 1000 times
2 : 100 times
1 : 10 times
0 : 1 times
○
5
D/A Zero Setting
000.00
-19999 to 19999
*
○
6
D/A Full Scale Setting
100.00
-19999 to 19999
*
○
7
D/A Output Mode
0
3 : 20mA (10V) fixed output
2 : 4mA (0V) fixed output
1 : Linked with indicated
value
0 : Not linked with indicated
value
○
8
Password
0000
1239 : Initialization
Set values can be rewritten to factory-set values
(
default
)
.
For simple initialization, see “Password” on page 99 , and for
simultaneous self-check and initialization, see “Initialization” on page 103 .
Содержание F320
Страница 1: ...F320 DIGITAL INDICATOR 17 May 2012 Rev 1 07 OPERATION MANUAL...
Страница 14: ...CONTENTS...
Страница 120: ...15 DIMENSIONS Standard 106 15 DIMENSIONS Standard Unit mm...
Страница 121: ...15 DIMENSIONS Standard 107 15 1 Dimensions When the BCO option is equipped Unit mm...
Страница 122: ...15 DIMENSIONS Standard 108 15 2 Dimensions When the DAV DAI option is equipped Unit mm...
Страница 129: ...Unipulse Corporation 9 11 Nihonbashi Hisamatsucho Chuo ku Tokyo 103 0005 Tel 81 3 3639 6120 Fax 81 3 3639 6130...