3.CONNECTION
15
3-2-2. DC Spec. ( Depending on the request at the time of order )
Connect the DC power cord. The input voltage is 12 to 24V (
±
15
%
) DC.
1. Make sure that no power is applied.
2. Remove the terminal block cover.
3. Remove the two screws at the lower left of the terminal block.
Connect the po) side of the power source to the red screw side of the
terminal block on the back of the F320, and its negative(-) side to the black
screw side.
Align the crimp contacts with the screw holes, and then tighten the screws.
4. Mount the terminal block cover.
B1
B2
B3
F.G.
(functional ground)
+
-
Within 6mm
Protective ground
The DC power code is not a standard accessory.
Use a source power (or battery) of 3A DC or more.