3.CONNECTION
10
3-1. Connecting Strain Gauge Sensor
Connect a strain gauge sensor.
The excitation voltage is selectable from 2.5 and 10V.
(The factory setting is 2.5V.)
The maximum output current is 30mA.
For the setting of the excitation voltage, see “Excitation Voltage” on page 65 .
◇
Attaching a Ferrite Core
Wind the sensor cable around the attached ferrite core.
If necessary fix the binding band.
B7
B8
B9
B10
Strain gauge sensor input
+EXC -SIG -EXC +SIG
Connect the shielded cable of the
strain gauge sensor to this screw.
Sensor cable
Wind the cable (one turn)
Ferrite core
F320
around the ferrite core.
Содержание F320
Страница 1: ...F320 DIGITAL INDICATOR 17 May 2012 Rev 1 07 OPERATION MANUAL...
Страница 14: ...CONTENTS...
Страница 120: ...15 DIMENSIONS Standard 106 15 DIMENSIONS Standard Unit mm...
Страница 121: ...15 DIMENSIONS Standard 107 15 1 Dimensions When the BCO option is equipped Unit mm...
Страница 122: ...15 DIMENSIONS Standard 108 15 2 Dimensions When the DAV DAI option is equipped Unit mm...
Страница 129: ...Unipulse Corporation 9 11 Nihonbashi Hisamatsucho Chuo ku Tokyo 103 0005 Tel 81 3 3639 6120 Fax 81 3 3639 6130...