Programming Manual MSO/UPO2000 Series
56 / 156
Instruments.uni-trend.com
:TRIGger:DURation:TIME:LOWer?
Query returns 1.000e000.
:TRIGger:DURation:TIME:UPPer
Command format
:
:TRIGger:DURation:TIME:UPPer <time>
:TRIGger:DURation:TIME:UPPer?
Functional description
:
This command is used to set the upper time limit of duration trigger. It can be set when time interval
condition is LESSthan (less than).
Return format
:
Query returns the current upper time limit, unit is s.
For example
:
:TRIGger:DURation:TIME:UPPer 1
Set the upper time limit of duration trigger to 1s.
:TRIGger:DURation:TIME:UPPer?
Query returns 1.000e000.
Setup Hold Trigger
:TRIGger:SHOLd:DATA:SOURce
Command format
:
:TRIGger:SHOLd:DATA:SOURce {CHANnel1 | CHANnel2| CHANnel3| CHANnel4}
:TRIGger:SHOLd:DATA:SOURce?
Functional description
:
This command is used to set data source of setup hold trigger.
Return format
:
Query returns {CHANnel1 | CHANnel2| CHANnel3| CHANnel4}.
For example
:
:TRIGger:SHOLd:DATA:SOUR CHAN1
Set channel 1 as data source.
:TRIGger:SHOLd:DATA:SOUR?
Query returns CHANnel1.
:TRIGger:SHOLd:CLOCK:SOURce
Command format
:
:TRIGger:SHOLd:CLOCK:SOURce {CHANnel1 | CHANnel2| CHANnel3| CHANnel4}
:TRIGger:SHOLd:CLOCK:SOURce?
Functional description
:
This command is used to set clock source of setup hold trigger.
Return format
:
Query returns {CHANnel1 | CHANnel2| CHANnel3| CHANnel4}.
For example
: