:SBUS:FR:PAYLoad:SIZe
Command format
:
:SBUS
:FR:PAYLoad:SIZe
<size>
:SBUS
:FR:PAYLoad:SIZe
?
Functional description
:
This command is used to set header field static load length in FlexRay bus trigger condition. Binary
character data presented by parameter 0, 1 or X. X represents uncertainty. Data bit is 7bits and its range is
0x0~0x7F.
Return format
:
Query returns binary character string data.
For example
:
:SBUS
:FR:PAYLoad:SIZe
"X00X00X1"
Set FlexRay bus static load length to X00X00X1.
:SBUS
:FR:PAYLoad:SIZe
?
Query returns X00X00X1.
:SBUS:FR:HEADer:CRC
Command format
:
:SBUS
:FR:HEADer:CRC
<crc>
:SBUS
:FR:HEADer:CRC
?
Functional description
:
This command is used to set header CRC in FlexRay bus trigger condition. Binary character data presented
by parameter 0, 1 or X. X represents uncertainty. Data bit is 11bits and its range is 0x0~0x7FF.
Return format
:
Query returns binary character string data.
For example
:
:SBUS
:FR:HEADer:CRC
"X00X00X1"
Set FlexRay bus header CRC to X00X00X1.
:SBUS
:FR:HEADer:CRC
?
Query returns X00X00X1.
:SBUS
:FR:CYCLes
Command format
:
:SBUS:FR:CYCLes <string>
:SBUS:FR:CYCLes?
Functional description
:
This command is used to set cycle number in FlexRay bus trigger condition. Binary character data
presented by parameter 0, 1 or X. X represents uncertainty. Data bit is 6bits and its range is 0x0~0x3F.
Return format
:
Query returns binary character string data.
For example
:
:SBUS:FR:CYCLes "0X00X1"
Set FlexRay bus cycle number to 0X00X1.
:SBUS:FR:CYCLes?
Query returns 0X00X1.