Programming Manual MSO/UPO2000 Series
25 / 156
Instruments.uni-trend.com
:FUNCtion:FFT:
DETEction
:AVERage:COUNt
Command format
:
:FUNCtion:FFT:DETEction
:AVERage:COUNt
<value>
:
FUNCtion:FFT:DETEction:AVERage
:COUNt
?
Functional description
:
This command is used to set the average number of times of average spectrum. The range is 2^n, n take
value from 1~10.
Return format
:
Query the average number of times of average spectrum.
For example
:
:FUNCtion:FFT:DETEction
:AVERage:COUNt 64
Set the average number of times of average spectrum to 64.
:FUNCtion:FFT:DETEction
:AVERage:COUNt
?
Query returns 64.
:
FUNCtion:FFT:DETEction:MAXHold
Command format
:
:FUNCtion:FFT:DETEction:MAXHold {OFF|PPEAK|NPEAK| AVERage|SAMPle}
:
FUNCtion:FFT:DETEction:MAXHold?
Functional description
:
This command is used to set the detection mode of the maximum hold spectrum.
OFF: Turn off the average spectrum;
PPEAK
:
Take the maximum value within the range of each sampling point;
NPEAK
:
Take the minimum value within the range of each sampling point;
AVERage
:
Take the average value within the range of each sampling point;
SAMPle
:
Take the value of first point within the range of each sampling point.
Return format
:
Query returns
the detection mode of the maximum hold spectrum.
For example
:
:FUNCtion:FFT:DETEction:MAXHold PPEAK
Set
the detection mode of the maximum hold spectrum to +peak detection.
:FUNCtion:FFT:DETEction:MAXHold?
Query returns PPEAK.
:
FUNCtion:FFT:DETEction:MINHold
Command format
:
:FUNCtion:FFT:DETEction:MINHold {OFF|PPEAK|NPEAK| AVERage|SAMPle}
:
FUNCtion:FFT:DETEction:MINHold?
Functional description
:
This command is used to set the detection mode of the minimum hold spectrum.
OFF: Turn off the average spectrum.
PPEAK
:
Take the minimum value within the range of each sampling point;
NPEAK
:
Take the minimum value within the range of each sampling point;