Programming Manual MSO/UPO2000 Series
22 / 156
Instruments.uni-trend.com
This command is used to select the
segment of FFT waterfall curve, which can select the range to 1~200.
Return format
:
Query returns which frame segment is currently selected.
For example
:
:
FUNCtion:FFT
:WATerfall:
SLICe
100
Set FFT waterfall curve to select 100 frame segment.
:
FUNCtion:FFT
:WATerfall:
SLICe
?
Query returns 100.
:FUNCtion:FFT:POINts
Command format
:
:
FUNCtion:FFT
:
POINts
{AUTO|1K|2K|4K|8K|16K|32K|64K|128K|256K|512K|1024K|2048K|4096K}
:
FUNCtion:FFT
:
POINts
?
Functional description
:
This command is used to set FFT’s point.
Return format
:
Query returns {AUTO|1K|2K|4K|8K|16K|32K|64K|128K|256K|512K|1024K|2048K|4096K}.
For example
:
:
FUNCtion:FFT
:
POINts 8K
Set FFT’s point to 8 K.
:
FUNCtion:FFT
:
POINts
?
Return 8 K.
:FUNCtion:FFT:VTYPe
Command format
:
:FUNCtion:FFT:VTYPe
{VRMS|DBRMS}
:FUNCtion:FFT:VTYPe?
Functional description
:
This command is used to select the unit of FFT vertical direction to dBRMS or VRMS. dBRMS represents
power root mean square, VRMS represents voltage root mean square.
Return format
:
Query returns VRMS, DBRMS.
For example
:
:FUNCtion:SOUR1 CHAN1
Set channel 1 as the source.
:FUNC:FFT:VTYP VRMS
Set the unit of FFT vertical direction to VRMS.
:FUNC:FFT:VTYP?
Query returns VRMS.
:FUNCtion:FFT:FREQuency
Command format
:
:FUNCtion:FFT:FREQuency?
Functional description
:
This command is used to acquire the center freuquency of spectrum waveform after FFT.
Return format
:
Query returns the center frequency of spectrum waveform, unit is Hz.
For example
:
:FUNCtion:FFT:FREQuency?
Query returns 1.000e003.