![UNI-T MSO/UPO2000 Series Скачать руководство пользователя страница 106](http://html1.mh-extra.com/html/uni-t/mso-upo2000-series/mso-upo2000-series_programming-manual_3695095106.webp)
Programming Manual MSO/UPO2000 Series
106 / 156
Instruments.uni-trend.com
:SBUS:FR:QUALifier?
Functional description
:
This command is used to set FlexRay bus trigger condition of the oscilloscope.
{START|IND|ID|CYCLES|HEAD|DATA|IDDATA|END|ERROR}: start frame, indicator bit, cycle number, header
filed, data, ID and data, end frame, error.
Return format
:
Query returns {START|IND|ID|CYCLES|HEAD|DATA|IDDATA|END|ERROR}.
For example
:
:SBUS:FR:QUALifier START
Set FlexRay bus trigger condition to start frame.
:SBUS:FR:QUALifier?
Query returns START.
:SBUS:FR:I
NDicator
Command format
:
:SBUS:FR
:I
NDicator {NORMal|STATIC|NULL|SYNC|START}
:SBUS:FR
:I
NDicator?
Functional description
:
This command is used to set FlexRay bus trigger indicator bit type of the oscilloscope.
{NORMal|STATIC|NULL|SYNC|START}: normal(01XX), static load (11XX), null (00XX), synchronization (XX10),
start (XX11).
Return format
:
Query returns {NORMal|STATIC|NULL|SYNC|START}.
For example
:
:SBUS:FR
:I
NDicator NORMal
Set FlexRay bus trigger indicator bit to normal.
:SBUS:FR
:I
NDicator?
Query returns NORMal.
:SBUS:FR:I
NDicator
:DATA
Command format
:
:SBUS
:FR:I
NDicator:DATA <string>
:SBUS
:FR:I
NDicator:DATA?
Functional description
:
This command is used to set header field indicator bit data in FlexRay bus trigger condition. Binary
character data presented by parameter 0, 1 or X. X represents uncertainty. Data bit is 5bits and its range is
0x0~0x1F.
Return format
:
Query returns character string type of 0x format.
For example
:
:SBUS
:FR:I
NDicator:DATA "000100X1"
Set FlexRay bus indicator bit data to 000100X1.
:SBUS
:FR:I
NDicator:DATA?
Query returns 000100X1.