Auto Regeneration Controller Model UK-2 Instruction Manual
4. DIP Switch Specifications
ions
There are two types of UK-2 DIP switch: Previous Model and New Model.
There are two types of UK-2 DIP switch: Previous Model and New Model.
Previous Model: Push a lever UP to turn ON. The color of the levers is yellow.
Previous Model: Push a lever UP to turn ON. The color of the levers is yellow.
(Previous models were sold until April, 2011.)
(Previous models were sold until April, 2011.)
N e w M o d e l : Push a lever DOWN to turn ON. The color of the levers is white.
N e w M o d e l : Push a lever DOWN to turn ON. The color of the levers is white.
(New model has been sold from May, 2011.)
(New model has been sold from May, 2011.)
Example: When No.9 and No.10 are ON.
Example: When No.9 and No.10 are ON.
[Previous Model] [New Model]
[Previous Model] [New Model]
Table 4-1
DIP Switch Descriptions
Table 4-1
DIP Switch Descriptions
(
※
1) No.9 and No.10 will not be used when serial communication is RS-232C.
(
※
1) No.9 and No.10 will not be used when serial communication is RS-232C.
No.
No. Item
Item
OFF
OFF
ON
ON
1
Prohibit panel control
Operable
Prohibited
2
Prohibit parameter change
Operable
Prohibited
3
BAND HEATER connection
Connected
Not connected
4
Cheesing Compressor type
C10series C15series C30series
4
OFF OFF OFF
5
ON ON OFF
5
6
ALARM Output
(SYSTEM I/F No.9)
Negative logic (b contact)
In normal status: ON
In ALARM status: OFF
Positive logic (a contact)
In normal status: OFF
In ALARM status: ON
7
INN HEATER connection
Not connected
Connected
8
Not in use
Do not turn ON
9
(
※
1)
Termination resistance
(
TXD
)
Not effective
Effective
(
RS485 only
)
10
(
※
1)
Termination resistance
(
RXD
)
Not effective
Effective
(
RS485 only
)
Do not make unnecessary change. Please contact us when the setting needs to be
changed.
CAUTION
When the specifications of cryopump and/or compressor which will be connected to UK-2
has been decided, DIP switch will be set before shipment.
ULVAC CRYOGENICS INCORPORATED
11
Содержание UK-2
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 10: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 46: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 48: ...Revision This page intentionally left blank R 2 ULVAC CRYOGENICS INCORPORATED ...