Auto Regeneration Controller Model UK-2 Instruction Manual
10
ULVAC CRYOGENICS INCORPORATED
Continued from previous page.
Table 3-5
Serial Communication RS I/F
Type
Pin
No.
Signal Name
Description
Input Signal
13 MODE
SELECT#1
Choosing MODE SELECT
#1 #2
Mode1 (FULL REGENE)
ON ON
Mode2 (WARM UP)
ON OFF
Mode3 (COOL DOWN)
OFF ON
14 MODE
SELECT#2
18 RESET
Reset the alarm output status. Turn this signal ON for
more than 0.1 second and turn it OFF.
15
〜
17
19
〜
25
N/C Do
not
connect.
※1
L level means that the transistor for open collector output is ON (establish electrical continuity).
H level means that the transistor for open collector output is OFF (break electrical continuity).
Terminal
No.
RS-485
(
Standard
)
RS-232C
(
Optional
)
1 TXD(+) TXD
2 TXD(-)
-
3 RXD(+) RXD
4 RXD(-)
-
5 FG FG
※
1 Standard UK-2 is equipped with RS-485.
Please contact our sales representative if RS-232C is preferred.
RS-232C can be installed instead of RS-485.
※
2 For the operation method by serial communication, refer to the instruction manual of
communication interface.
RS
I/F
ID
0
1
2
3
4
5 6
7
8
9
1
2
3
4
5
Содержание UK-2
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 10: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 46: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 48: ...Revision This page intentionally left blank R 2 ULVAC CRYOGENICS INCORPORATED ...