Auto Regeneration Controller Model UK-2 Instruction Manual
22
ULVAC CRYOGENICS INCORPORATED
8. Operation
8.1 Remote Operation
①
Turn ON the [REMOTE] signal of SYSTEM I/F.
Ensure that key LED on the front panel is lit.
②
Determine the [MODE SELECT #1] and [MODE SELECT #2] signals of SYSTEM I/F.
MODE SELECT #1
MODE SELECT #2
Mode1 (FULL REGEN) ON
ON
Mode2 (WARM UP)
ON
OFF
Mode3 (COOL DOWN) OFF
ON
Cancel
OFF
OFF
③
Turn ON the [REGEN] signal of SYSTEM I/F.
It starts operating in the mode that has been selected above.
8.2 Local Operation
①
Ensure that key LED is light off.
If the LED is lit, hold down the key for 1 second and turn the light off.
→
It will move to manual operation mode.
②
Hold down key for 1 second.
→
key LED lights up.
③
Use
keys to select an operation mode referring the table below.
→
LEDs flash as described below in each mode.
Operation Mode
Mode 1
Mode 2
Mode 3
None
Flashing LED
HEATER START
WARM UP
BUILD UP
ROUGHING
N2 PURGE
COOL DOWN
20K(2ND)
HIGH VAC OK
HEATER
START
WARM UP
BUILD UP
ROUGHING
N2 PURGE
COOL DOWN
20K(2ND)
HIGH VAC OK
HEATER
START
WARM UP
BUILD UP
ROUGHING
N2
PURGE
COOL DOWN
20K(2ND)
HIGH VAC
OK
HEATER START
WARM UP
BUILD UP
ROUGHING
N2 PURGE
COOL DOWN
20K(2ND)
HIGH VAC
OK
Operation
Operate warm
up to cooldown.
Operate warm
up to rough
pumping.
Operate
cooldown only.
Cancel
④
After selecting the mode, press key to start the operation.
⑤
To stop the operation, hold down key for 1 second.
REMOTE
REMOTE
REGEN
REGEN
REGEN
REMOTE
REGEN
Содержание UK-2
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 10: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 46: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 48: ...Revision This page intentionally left blank R 2 ULVAC CRYOGENICS INCORPORATED ...