Auto Regeneration Controller Model UK-2 Instruction Manual
6
ULVAC CRYOGENICS INCORPORATED
3.2 Rear Panel Description
Table 3-2
Part Names and Descriptions
(
Rear Panel
)
Table 3-2
Part Names and Descriptions
(
Rear Panel
)
No.
No.
Name
Name
Description
Description
①
SOURCE POWER Main circuit protector of UK-2.
(
5A
)
②
HEATER POWER Circuit protector of BAND HEATER. (15A)
③
SOURCE INPUT
Power input of UK-2. AC100
〜
220V±10%
(Power input must be AC100V ±10% when using INNER HEATER)
To connect our INNER HEATER (1st, 2nd HEATER), always input
AC100V to the SOURCE INPUT or it may result in breakage of the
HEATER.
(
SOURCE INPUT double as INNER HEATER output power source
and the capacity of the INNER HEATER is AC100V.
)
④
BAND HEATER
BAND HEATER input and output (AC200
〜
220V±10% MAX15A).
⑤
SENSOR INPUT
1ST TEMP and 2ND TEMP
,
PIRANI GAUGE input.
⑥
ATM
DC24V output (For MBS-C, PIRANI GAUGE input power), ATM input.
⑦
SYSTEM I/F
I/F for remote operation from the equipment.
⑧
COMP I/F
I/F for the Compressor.
⑨
VALVE
N2 IN/OUT VALVE, ROUGH VALVE control output.
⑩
HEATER
Power output of INNER HEATER.
(Output the same voltage as it has input to SOURCE INPUT.)
⑪
SW
Function choosing Dip SW
(
See section.4
)
13
12
11
10
9
8
7
6
5
4
3
2
1
910
1 2 34 5 67 8
HEATER
POWER
SOURCE
POWER
HEATER
S
R
N/C
V
U
N/C
N/C
N/C
RS I/F
BAND
HEAT
ER
A
C
220
V
15A
MAX
SW
ATM
ID
VALVE
COMP
I/F
SENSOR
INPUT
SYSTEM
I/F
SOURCE
INPUT
CAUTION
Содержание UK-2
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 10: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 46: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 48: ...Revision This page intentionally left blank R 2 ULVAC CRYOGENICS INCORPORATED ...