Auto Regeneration Controller Model UK-2 Instruction Manual
Continued from previous page.
Display Class
Description
Possible Cause
Corrective Action
_cP
A
Alarm output from
compressor.
Compressor
abnormal.
Please refer to the
compressor manual.
A
When there is no
COMP ON ANS (for
5 seconds) during
COMP ON, alarm
will be activated.
_ht
A
After cooldown has been completed
(
HI VAC OK
)
, it will be displayed if
1ST TEMP exceeds 1AL (150K) or 2ND TEMP exceeds 2AL (20K).
oh
A
Alarm will be output
when the
temperature
exceeds oh set
temperature (340K).
Heat input from heater
and/or other heat
sources.
Cut off all heat sources
immediately.
ALARM
CLASS Description
S1:
Requires repair
System stop
Requires adjustment service
S2:
Requires resetting
System stop
Requires user setting
A:
Severe failure
Stops full output
Outputs ALARM
B:
Minor failure
Continues outputting Outputs ALARM
Table 10-2
Troubleshooting (other errors)
Display
Class
Description
Corrective Action
cPu1
cPu2
S1
EEPROM read error, ROM versionerror.
EEPROM adjustment area check SUM
error.
Please contact our
customer service.
※
Error code cannot
be reset.
EPEr
S2
EEPROM parameter area check
SUMerror.
Go back to default
setting by resetting
error alarm.
30
ULVAC CRYOGENICS INCORPORATED
Содержание UK-2
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 10: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 46: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 48: ...Revision This page intentionally left blank R 2 ULVAC CRYOGENICS INCORPORATED ...