Auto Regeneration Controller Model UK-2 Instruction Manual
※
1
It will not displayed when DIP SW
⑦
is OFF (means inner heater is not used).
※
2
Heater maximum energizing time
When heater is energized more than the set value, an alarm will be activated.
(
Recommended set values
)
Pump size
Set time
8 inch
45 min
+
Pgt
12 inch
80 min
+
Pgt
16 inch (U16P)
150 min
+
Pgt
16 inch other than above and larger
size
(U16HSP, U16HSPL)
240 min
+
Pgt
28
ULVAC CRYOGENICS INCORPORATED