Auto Regeneration Controller Model UK-2 Instruction Manual
7. Wiring
After installing UK-2 to the panel, perform wiring referring to Figure 7-1. Purchased
optional cables must be connected as shown in Figure 7-2. Observe following warnings
and cautions to perform wiring.
WARNING
●
Always disconnect this product from any power source during wiring operation to prevent
electrical shock.
●
Avoid touching the wired terminal and charged devices while supplying power.
CAUTION
●
ANALOG signal and
K
Thermocouple
①
!
For ANALOG signal and K Thermocouple input signal, use cables with shield.
Especially for K thermocouple, use the compensation lead wire with shield which is
effective for electric noise.
②
!
Ensure to wire MBS-C POWER properly. If it is not wired correctly, it may result in the
breakage of MBS-C.
CAUTION
●
Conduit Requirements for UK-2 wiring
To prevent interference during the
operation, there should be a separate conduit for each
signal line, control line, and AC power line. Especially the ANALOG signal line of MBS-C
must be in separate conduit from other equipments or AC power lines or control lines for
other equipments. Otherwise, UK-2 operation will be interfered.
If separate conduits cannot be supplied, keep enough distance (generally 300mm or more)
between different wirings. It is effective to prevent interference.
16
ULVAC CRYOGENICS INCORPORATED
Содержание UK-2
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 10: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 46: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 48: ...Revision This page intentionally left blank R 2 ULVAC CRYOGENICS INCORPORATED ...