Page 29
9.4.
Maintenance
Reset Atmospheric Pressure
1.
This subdirectory allows to reset the atmospheric pressure.
2.
On the Main Menu, choose Maintenance\ Reset Atmospheric Pressure.
3.
The following message appears on the screen: "Reset done! New value will be
set after door is open for 2 minutes and the temperature is less than 45.0 °C".
10.
MAINTENANCE INSTRUCTIONS
10.1. Preventive and Scheduled Maintenance
The maintenance operations described in this chapter need to be followed as indicated to
keep the device in good working condition.
The instructions that follow can easily be carried out by the operating personnel and do not
require a service technician.
Note:
Technician manual describes the maintenance operations required from qualified technician,
every two months and once a year.
Should the need arise, technical assistance or a service technician can be requested by
calling your dealer.
Daily
Clean door gasket with a mild detergent, water and a soft cloth or
sponge. The gasket should be clean and smooth.
.
Weekly by the operator
Empty and refill the vacuum pump reservoir.
Put a few oil drops on the 2 door pins and door tightening bolt.
Clean the outer parts of the autoclave with a soft cloth.
Clean the chamber.
Periodically by the operator
Once per month clean the water outlet strainer (see 10.3). Cleaning
frequency may be reduced according to experience.
Check the door gasket every 12 months and replace it if required (see
Technician manual).
Replace the air filter, every 6 months or after 1000 cycles (whichever
comes first) according to sec. 12.3
.
10.2. Cleaning the Water Outlet Strainer
Cautions!
Before proceeding, make sure that the electric cord is disconnected and
there is no pressure or water in the chamber.
Warnings
1. The strainer’s cover is HOT
Do not touch the strainer’s cap, mounted on the exhaust line,
during and shortly after operation. Touching the hot strainer’s
cap may cause severe injuries.
2. If maintenance operation is performed while strainer cap is hot,
use heat resistant gloves to avoid injuries.
1.
Open the strainer cap.
2.
Remove the strainer element.
3.
Rinse the strainer with water, using a brush if necessary.
4.
Reinstall the strainer element.
5.
Close the strainer cap.