Page 8
Cautions!
In order to avoid any injury by electrical hazard, it is recommended that
a ground fault protection device (GFCI) be installed in the electrical
panel feeding the autoclave (local codes may make this mandatory).
The electrical network must comply with local rules or regulations.
Verify that there is an easy access to the main power switch and to the
current leakage safety relay (GFCI). The voltage supplied to the device
must comply within the label value± 5%.
3.7.
Environmental Emission Information
1.
The peak sound level generated by the autoclave is 70dBa with background
noise of 60dBa.
2.
The total heat per hour transmitted by the autoclave is <300Wh.
3.8.
Construction
The main parts of the autoclave are made of materials as indicated below:
Chamber is built of stainless steel 316 L.
Door is made of stainless steel 304.
.
Cabinet is made of stainless steel 304
Trays are made of stainless steel 316L.
Generator's vessel is made of steel 285C
Door closing handle is made of aluminium material, which is safe to
touch and thermo-insulated.
3.9.
Stickers Description
Symbol
Meaning
Part Number
Location
Caution! Consult
accompanying
documents
LAB048-0024
Near the cut-off
Caution! Hot steam.
LAB048-0058
Near the safety valve
Protective earth
(Ground)
LAB048-0020
Near the on/off switch
ON
OFF
On-Off
LAB048-0018
Near the Power switch