Page 18
electric shock, which may lead to fires or accidents
.
Cautions!
Never use the printer in a place of extreme humidity or any place where
it can possibly be splashed by any liquids. If any liquids get into the
printer, it could lead to fire, electric shock, or other serious accidents
.
Never touch the thermal head immediately after printing because it
becomes very hot. Make sure that the thermal head is cool before
setting papers or cleaning the thermal head.
Power OFF the autoclave in any of the following cases:
The printer does not recover from an error.
Smoke, strange noise or smells erupt from the printer.
A piece of metal or any liquid touches the internal parts or slot of the
printer.
4.2.
Displayed Error Messages / Symbols
The failures are divided into two categories:
Failures that occur before completing the sterilization stage, which in
this case will leave the load unsterilized
Failures that occur after completing the sterilization stage, which in this
case will leave the load sterilized
For the list of
Displayed Error Messages / Symbols
see 11 TROUBLESHOOTING
4.3.
Displayed Operational Messages/ Symbols
Message /
Symbol Name
Message / Symbol Description
Required Action
This symbol is displayed when the door
is open.
Close the door.
"Door is open"
This message is displayed when the
door is opened: In stand–by, if start key
is pressed.
Close the door to perform a
new cycle.
If the problem persists, call the
technician.
"Cycle Ended"
This message is displayed when the
cycle ended successfully.
Press START/STOP in order to
perform a new cycle.
“Test Ended“
This message is displayed when the
test ended.
Press START/STOP in order to
perform a new test
This symbol is displayed when Cycle by
Clock mode is active.
Enter the main menu as
described in this manual to
change the time or to cancel
this option.
This symbol is displayed when the
temperature in the jacket did not reach
the preset value
Wait until the temperature in
the jacket reaches the preset
value
"System is not
ready to run a
cycle.Cycle by
clock is active”
This message is displayed if the user
presses START/STOP key while the
"start cycle by clock" mode is active.
Starting another cycle is not allowed.
Enter the main menu as
described in this manual to
change the time or to cancel
this option.
"Atmospheric
pressure not set"
This message is displayed in order to
set the atmospheric pressure by
opening the door for 2 minutes.
Open the door for 2 minutes in
order to set the Atmospheric
pressure.
"Please restart
machine in order
for changes to be
updated"
Changes to the system software
require that the autoclave be restarted.
Restart the autoclave in order
for changes to be updated.