Page 21
5.4.
Bowie and Dick Test Process Description
Air-removal stage: vacuum pulses are performed.
Heating stage: steam is inserted into the chamber until the sterilization
temperature and pressure are reached.
Sterilization stage: temperature and pressure are maintained constant
at the pre-set level for sterilization time
.
Fast exhaust stage; steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure
.
Drying stage; heating of chamber followed by a vacuum break (air inlet)
to remove leftover moisture from the instruments and wraps
.
Air inlet to
reach atmospheric pressure.
Notes:
If the B&D test cycle failed, the screen color changed from purple to yellow
If the B&D test cycle ended successfully "Test Ended" message will display and the
screen color will remain purple.
6.
INSTALLATION INSTRUCTION
6.1.
Placing
Make sure while placing the autoclave, to leave space around the machine, to give the
technician access to service the machine. Keep the back and the right side of the autoclave
approximately 2” (50mm) away from the wall to allow for ventilation.
At the time of installation or anytime the unit is relocated the atmospheric pressure parameter
needs to be reset. (See 9.4).
Connect the power cord to the socket on the rear side of the autoclave; plug it into the supply
outlet. The autoclave must be connected to a properly grounded outlet.
6.2.
Setup
Make sure all the feet are on the autoclave and none of them has been lost
The unit is ready to operate.
6.3.
Utilities connection
If the autoclave is intended to operate with water inlet and drain outlet
connected to the facility's utilities, connect the following:
a. Connect the autoclave to the source of mineral-free water through a
flexible ½” hose. Mount a manual valve on the source of the water
supply, in order to enable closing the water inlet for maintenance or
servicing purposes.
b. Connect the autoclave's drain to the building’s drainage pipe. The
drainage shall be of an open type, withstanding temperature of, at
least,
60
º
C. If you wish to use the autoclave without drain
connection, put a bucket under the feed water automatic drain
.
c. Connect the feed water supply to the tap water inlet. The supplied
water pressure shall be set to 2-4bar. Mount a manual valve on the
source of the water supply, in order to enable closing the water inlet
for maintenance or servicing purposes.
6.4.
Moving the unit
Caution!
Before moving the autoclave, make sure that the electric cord is
disconnected from the power, and there is no pressure in the chamber.
1. Disconnect the power supply cord.
2. Drain the water from the reservoir and the chamber (open the water
drain valve (see Rear View) and pour the water to a bucket)