Page 27
during and shortly after operation. It will get very hot.
Touching the hot strainer’s cover may cause severe injuries.
To avoid severe injuries from hot steam when opening the
door, it is strictly forbidden to lean on the autoclave or to
place your hand or any part of your body over the door.
8.6.
Stopping the Process Manually
It is possible to stop the program while the autoclave is operating. Pressing the
START/STOP key at any stage of the process stops the operation.
If the cycle was aborted before completing the sterilization stage, it will leave the load
unsterilized. If the cycle is aborted before completing the sterilization stage, a yellow
caution screen is displayed with a caution symbol, the message "Cycle Failed", and an
error message explaining the reason for the failure.
Pressing the START/STOP key cancels the displayed message and unlocks the door so
it can be opened.
Warnings:
The sterility of the instruments processed in unwrapped
cycles cannot be maintained if exposed to non-sterile
environment.
Don’t touch the strainer’s cover, mounted on the exhaust line,
during and shortly after operation. It will get very hot.
Touching the hot strainer’s cover may cause severe injuries.
To avoid severe injuries from hot steam when opening the
door, it is strictly forbidden to lean on the autoclave or to
place your hand or any part of your body over the door.
The load has not completed a sterilization cycle, therefore it is not sterile.
Handle it as contaminated load.
8.7.
Stopping the Process due to Cycle Failure
The cycle can stop itself if the unit detects a problem
If the cycle is aborted prior to completing the sterilization stage, a yellow
caution screen is displayed with a caution symbol, the message "Cycle
Failed", and an error message explaining the reason for the failure.
Pressing the START/STOP key cancels the displayed message and
unlocks the door so it can be opened.
Warning:
The load has not completed a sterilization cycle, therefore it is not sterile.
Handle it as contaminated load.
9.
CHECKING AND CHANGING PARAMETERS AND OTHER DATA
9.1.
Quick Options Screen
1.
When the system is ready, enter the QUICK OPTIONS screen by pressing
the UP and DOWN keys simultaneously.
Note:
To exit every screen do one of the following:
move the cursor to Exit by pressing UP or DOWN keys and then press
START/STOP key.
press the UP and DOWN keys simultaneously.
On this screen you can either proceed to login (see 9.2) or choose one of the
quick options available without login.