Page 25
7.3.
Tubing
Rinse tubing after cleaning with pyrogen free water. When placing in a
tray, Make sure that both ends are open, without sharp bends or twists.
7.4.
Wrapped Instruments
1.
Wrapped instruments should be packed in material that promotes drying such
as autoclave bag, autoclave paper.
2.
Use single-use wraps once only and discard after use.
3.
It is highly recommended to utilize the Tuttnauer
Pouch Rack. This rack
allows the operator to place pouches on their side, thus increasing the
capacity of the autoclave significantly and promoting better drying of the
instruments. Contact your dealer for details.
4.
Verify that the packaging method is in accordance with good practice
approach and the packaging materials are in accordance with the applicable
standards. We recommend using plastic-paper pouches.
8.
OPERATING INSTRUCTIONS
8.1.
Water Supply
1.
If the autoclave is connected to external water supply, open the manual
valves of the tap water supply and mineral-free water supply.
8.2.
Turning on the Autoclave
1.
Plug the power cord into the back of the autoclave and into the wall outlet.
2.
Turn on the rocker switch (1), located behind the printer's door.
3.
When running the autoclave for the first time, set date and time (see , 9.1
Quick Options Screen, Set Date and Time).
8.3.
Opening the door – HSG
1.
Move the door handle (1) from the position "locked" (2) to the position
"unlocked" (3).
2.
Open the door.
1
Wrong
Right